Leaflet Modification Technique With UNICORN in Failed Aortic Bioprosthesis: Procedural Step-by-Step, Best Practice, and Troubleshooting.

📚 期刊: JACC. Cardiovascular interventions 📅 发表: 0000-00-00 🔬 PMID: 42331405 🔗 DOI: 10.1016/j.jcin.2026.04.031 👁️ 浏览: 1

👤 作者: Chan KE, Un KC, Chiu HAC, Tai-Leung Chan D, Sze SY, Yin WH, Lee YT, Lee YT, Lin P, Abdel-Wahab M

心脏瓣膜

📝 摘要

The UNICORN (undermining iatrogenic coronary obstruction with radiofrequency needle) technique was developed in 2022 to reduce the risk for coronary obstruction during valve-in-valve transcatheter aortic valve replacement. The procedure involves traversing the target leaflet with an electrified wire, followed by serial dilation with coronary and peripheral balloons. A balloon-expandable valve is then delivered into the de novo fenestration for intraleaflet deployment. As the balloon-expandable valve expands, the target leaflet is opened and displaced away from the at-risk coronary ostium, resulting in a minimal neoskirt on the target side. Since its introduction, the technique has been adopted at numerous centers worldwide, with various modifications and adaptations. In this paper, a step-by-step description of the UNICORN procedure is provided, and troubleshooting strategies for potential challenges are discussed.